Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 30: (1 year) CAL VER /5520 DATE: 18-Feb-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 104 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # Fluke Model 30 Clamp Meter Service Information (PN 929042 January 1994). # Fluke 5520A/COIL Instruction Sheet (PN 600616 Aug. 1995) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 DISP {System accuracy of 5520A + actual 50-turn} 1.002 DISP {toroid used must be entered for each ACC statement} 1.003 DISP {to reflect actual system accuracy. } 1.004 DISP {See procedure README file.} 1.005 ASK- P F W 1.006 ASK+ K 1.007 HEAD EQUIPMENT SETUP 1.008 DISP This procedure requires a 50-turn Toroid Coil. 1.008 DISP Refer to the Model 30 Clamp Meter Service Information 1.008 DISP for directions for constructing a suitable coil. 1.009 HEAD {CURRENT} 1.010 DISP Connect the 5520A NORMAL Hi & Lo to the UUT V Ohm & COM 1.010 DISP terminals. 1.010 DISP 1.010 DISP Connect the 5500A/COIL to the 5520A AUX Hi & Lo 1.010 DISP terminals. 1.010 DISP 1.010 DISP Clamp the meter around the toroid center conductor 1.010 DISP bundle, positioning at the alignment marks. 1.011 HEAD CURRENT: {200A Range} 1.012 5520 20.0A 60H SI S LT5 # System accuracy of 5520A + actual 50-turn toroid used must be entered here. 1.013 ACC 20.A 1.2U 1.014 MEMI Enter UUT reading in Amps: 1.015 MEME 1.016 MEMC 200 A 1.2U 60H 2.001 DISP Connect the 5500A/COIL to the 5520A 20A & AUX Lo 2.001 DISP terminals. 2.002 5520 190.0A 60H SI HC S LT5 # System accuracy of 5520A + actual 50-turn toroid used must be entered here. 2.003 ACC 190.0A 2.7U 2.004 MEMI Enter UUT reading in Amps: 2.005 MEME 2.006 MEMC 200 A 2.7U 60H 3.001 HEAD CURRENT: {400A Range} 3.002 DISP Connect the 5500A/COIL to the 5520A AUX Hi & Lo 3.002 DISP terminals. 3.003 5520 40A 60H SI S LT5 # System accuracy of 5520A + actual 50-turn toroid used must be entered here. 3.004 ACC 40A 4U 3.005 MEMI Enter UUT reading in Amps: 3.006 MEME 3.007 MEMC 400 A 4U 60H 4.001 DISP Connect the 5500A/COIL to the 5520A 20A & AUX Lo 4.001 DISP terminals. 4.002 5520 350A 50H SI HC S LT5 # System accuracy of 5520A + actual 50-turn toroid used must be entered here. 4.003 ACC 40A 8U 4.004 MEMI Enter UUT reading in Amps: 4.005 MEME 4.006 MEMC 400 A 8U 50H 5.001 5520 400A 50H SI HC S LT5 # System accuracy of 5520A + actual 50-turn toroid must be entered here. 5.002 ACC 40A 12U 5.003 MEMI Enter UUT reading in Amps: 5.004 MEME 5.005 MEMC 400 A 12U 50H 6.001 HEAD {VOLTAGE} 6.002 HEAD VOLTAGE: {200V Range} 6.003 5520 200 20.0V 0.5U 60H SI 2W 7.001 5520 200 190.0V 2.6U 60H SI 2W 8.001 HEAD VOLTAGE: {600V Range} 8.002 5520 600 60V 4U 60H SI 2W 9.001 5520 600 600V 10U 60H SI 2W 10.001 HEAD {RESISTANCE} 10.002 5520 200 190Z 3.1U 2W 11.001 END